A Novel Capacitive Method for Urodynamic Measurements That is Suitable for Home Use

Nova metoda merjenja urodinamike, primerna za domačo uporabo na osnovi kapacitivnosti

Authors

  • Gregor Nikolić University of Maribor, Faculty of Electrical Engineering and Computer Science, Laboratory for Electronic and Information Systems , Univerza v Mariboru, Fakulteta za elektrotehniko, računalništvo in informatiko, Laboratorij za elektronske in informacijske sisteme Author
  • Igor But University Medical Center Maribor, Division of Gynecology and Perinatology, Department of General Gynecology and Gynecologic Urology , Univerzitetni klinični center Maribor, Klinika za ginekologijo in perinatologijo, Oddelek za splošno ginekologijo in ginekološko urologijo Author
  • Andraž Stožer University of Maribor, Faculty of Medicine, Institute of Physiology , Univerza v Mariboru, Medicinska fakulteta, Inštitut za fiziologijo Author
  • Iztok Kramberger University of Maribor, Faculty of Electrical Engineering and Computer Science, Laboratory for Electronic and Information Systems , Univerza v Mariboru, Fakulteta za elektrotehniko, računalništvo in informatiko, Laboratorij za elektronske in informacijske sisteme Author

DOI:

https://doi.org/10.18690/actabiomed.195

Keywords:

urodinamske meritve, urodinamika, neinvazivne meritve urodinamike, simptomi spodnjih sečil, disfunkcija uriniranja, prekomerno dejaven mehur

Abstract

Purpose: Urinary incontinence is highly prevalent in both women and men. Urodynamic testing is essential for diagnosis, but a gap exists in the usefulness of simple assessments that do not provide sufficient data, and devices that provide the necessary data, but are used primarily in clinics and clinical trials. This study considered the existing methods of urodynamics testing and developed a device that was both effective and suitable for home use.

Methods: A review of existing urodynamic methods led to development of a capacitive device. We created a device with sensor electrodes placed on the outside of a specimen container. The functionality of the electrodes and performance of the novel capacitive device were compared with that of a gravimetric device (Dorado KT, Laborie Medical Technologies, Mississauga, ON, Canada) used at the University Clinical Center, Maribor.

Results: Measurements of fluid volume (p = 0.936), maximum (p = 0.327), and mean (p = 0.136) flow observed with the two devices were not significantly different. Graphical analysis of collected volumes and flow variation over time show a high degree of overlap.

Conclusion: Compared with previously developed urodynamics testing devices, the capacitance method and the performance of this device represent an advance. The precision of the measurements needs improvement and did not fully address the recommendations of the International Continence Society, but he use of additional sensor electrodes is expected to improve performance. Extending the functionality as a standalone capacitive device for home use and upgrading with wireless access by mobile systems would allow for additional benefits beyond current voiding diaries.

Author Biographies

  • Gregor Nikolić, University of Maribor, Faculty of Electrical Engineering and Computer Science, Laboratory for Electronic and Information Systems, Univerza v Mariboru, Fakulteta za elektrotehniko, računalništvo in informatiko, Laboratorij za elektronske in informacijske sisteme

    Maribor, Slovenia.

  • Igor But, University Medical Center Maribor, Division of Gynecology and Perinatology, Department of General Gynecology and Gynecologic Urology, Univerzitetni klinični center Maribor, Klinika za ginekologijo in perinatologijo, Oddelek za splošno ginekologijo in ginekološko urologijo

    Maribor, Slovenia.

  • Andraž Stožer, University of Maribor, Faculty of Medicine, Institute of Physiology, Univerza v Mariboru, Medicinska fakulteta, Inštitut za fiziologijo

    Ph. D., Maribor, Slovenia. E-mail: andraz.stozer@um.si

  • Iztok Kramberger, University of Maribor, Faculty of Electrical Engineering and Computer Science, Laboratory for Electronic and Information Systems, Univerza v Mariboru, Fakulteta za elektrotehniko, računalništvo in informatiko, Laboratorij za elektronske in informacijske sisteme

    Ph. D., Maribor, Slovenia. E-mail: iztok.kramberger@um.si

Downloads

Published

29.11.2021

Issue

Section

Laboratory Study

How to Cite

Nikolić, G., But, I., Stožer, A. ., & Kramberger, I. . (2021). A Novel Capacitive Method for Urodynamic Measurements That is Suitable for Home Use: Nova metoda merjenja urodinamike, primerna za domačo uporabo na osnovi kapacitivnosti. Acta Medico-Biotechnica, 13(1), 50-58. https://doi.org/10.18690/actabiomed.195

Most read articles by the same author(s)