Peak Torque Comparison between SMM iMoment and Biodex System Pro 4 Isokinetic Dynamometers

Primerjava največjega navora med izokinetičnima dinamometroma SMM iMoment in Biodex System Pro 4

Authors

  • Matjaž Vogrin University of Maribor, Faculty of Medicine, Institute of Sports Medicine; UniversityMedical Centre Maribor, Division of Surgery, Department of Orthopaedics; University of Maribor, Faculty of Medicine, Department of Orthopedics , Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino; Univerzitetni klinični center Maribor, Klinika za kirurgijo, Oddelek za ortopedijo; Univerza v Mariboru, Medicinska fakulteta, Katedra za ortopedijo Author
  • Teja Ličen University of Maribor, Faculty of Medicine, Institute of Sports Medicine , Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino Author
  • Nina Greiner University of Maribor, Faculty of Medicine, Institute of Sports Medicine Maribor , Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino Author
  • Miloš Kalc University of Maribor, Faculty of Medicine, Institute of Sports Medicine , Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino Author

DOI:

https://doi.org/10.18690/actabiomed.204

Keywords:

inter-machine reliability, knee flexors, angle of peak torque, peak torque

Abstract

Purpose: To compare peak torque of knee extensors and flexors during muscle contractions between the SMM iMoment and Biodex System Pro 4 isokinetic dynamometers (inter-machine reliability).
Methods: Twenty-five students (14 men and 11 women, age 24.35 (1.41) years) took part in a crossover study. Peak torque (PT) and angle of peak torque (APT) for knee extensors and flexors were assessed at the velocity of 60°/s and 180°/s on two visits. Paired t-test, intraclass correlation coefficient (ICC), standard error of measurement (SEM), coefficient of variation (CVmed) and Bland-Altman plots were used to determine significant differences, relative and absolute reliability and agreement between devices.
Results: There were statistically significant differences in PT between machines at the velocity of 60°/s for the right leg knee flexion (p = 0.027). ICC values revealed high to very high relative reliability for the left leg (0.84 - 0.93) and moderate to very high for the right leg (0.79 - 0.91). Values of absolute reliability revealed more substantial differences between dynamometers compared to other similar studies (CVmed: 4.84% - 9.36%, SEM: 9.62 Nm - 26.11 Nm). Bland-Altman plots revealed no specific bias in exercise performed with the left leg and small bias (4-6 Nm) in exercises performed with the right leg. There were significant differences in APT between dynamometers in all conditions.
Conclusion: Mean peak torque values of knee extensors and flexors were similar between machines; however, data collected using the iMoment dynamometer were substantially more variable. 

Author Biographies

  • Matjaž Vogrin, University of Maribor, Faculty of Medicine, Institute of Sports Medicine; UniversityMedical Centre Maribor, Division of Surgery, Department of Orthopaedics; University of Maribor, Faculty of Medicine, Department of Orthopedics, Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino; Univerzitetni klinični center Maribor, Klinika za kirurgijo, Oddelek za ortopedijo; Univerza v Mariboru, Medicinska fakulteta, Katedra za ortopedijo

    Maribor, Slovenia.

  • Teja Ličen, University of Maribor, Faculty of Medicine, Institute of Sports Medicine, Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino

    Maribor, Slovenia.

  • Nina Greiner, University of Maribor, Faculty of Medicine, Institute of Sports Medicine Maribor, Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino

    Maribor, Slovenia.

  • Miloš Kalc, University of Maribor, Faculty of Medicine, Institute of Sports Medicine , Univerza v Mariboru, Medicinska fakulteta, Inštitut za športno medicino

    Maribor, Slovenia. E-mail: milos.kalc@ism-mb.si (Correspondence)

Downloads

Published

25.11.2021

Issue

Section

Laboratory Study

How to Cite

Vogrin, M., Ličen, T., Greiner, N., & Kalc, M. (2021). Peak Torque Comparison between SMM iMoment and Biodex System Pro 4 Isokinetic Dynamometers: Primerjava največjega navora med izokinetičnima dinamometroma SMM iMoment in Biodex System Pro 4. Acta Medico-Biotechnica, 13(2), 46-54. https://doi.org/10.18690/actabiomed.204

Most read articles by the same author(s)